21
Reference Manual
00809-0100-4811, Rev DA
Section 3: Installation
January 2015
Installation
Figure 3-1. Switches for Alarm and Write Protection
To set the alarm and write protect switches do the following:
1. Remove the cover on the circuit side (see main label).
2. To set the 4-20 mA alarm output to Low, move the alarm switch to the LOW position.
The factory default setting is HIGH (see Figure 3-1).
3. To enable the security write protection feature, move the write protect switch to the
ON position. The factory default setting is the OFF position (see Figure 3-1).
4. Replace and tighten the cover.
Alarm output
Write protection
Vedere la pagina 30