Reference Manual
00809-0100-4811, Rev DA
Appendix C: Rosemount 3300 Series with HART
®
to Modbus
®
Converter
January 2015
Rosemount 3300 Series with HART
®
to Modbus
®
Converter
Table C-7. Implemented Functions of Levelmaster Protocol
Table C-8. Letters and Expressions Used in Previous Tables
C.8.5 Modbus ASCII communication setup
The parameter, default, and configurable values are shown in Ta bl e C -9 below.
Table C-9. Modbus ASCII Communication Parameters
Input format Description Output format
UnnN? Return ID number UnnNnnCcccc
UnnNmm Set ID number UnnNOKCcccc
UnnF? Return number of floats UnnFxCcccc
UnnFx? Set number of floats UnnFOKCcccc
Unn? Return floats and other data UnnDddd.ddFfffEeeeeWwwwCcccc
(1)
(1) In this case, number of floats is set to 1. If number of floats is set to 2, the Output Format would be:
UnnDddd.ddDddd.ddFfffEeeeeWwwwCcccc
Note
If one float is sent, it is “Float1”. If two floats are sent, it is “Float 1” before “Float 0”.
Letter Description
nn
nn is used to identify slave to respond,
nn is a number 00-99 or ** (wildcard).
The EmulCtrl Address Holding register can be configured to a higher value than 99. In that
case, the address will be truncated to 99.
mm mm is the new ID number for the slave; mm is a number 00-99.
x x is the number of floats returned when slave receives Unn?, x is a number 0-2.
cccc Is the 16 bit CRC checksum, cccc are hexadecimal characters.
ddd.dd ddd.dd is the distance value from slave 1. Note that the first d can also be a ‘-’ (minus).
Float 1 Slave 1 PV.
Float 0 Slave 1 SV.
fff The temperature value. Configured by Holding Register 3208 in HMC.
(1)
(1) Any of the four available variables from any of the five HART slaves can be selected as the temperature source.
The least four significant bits (bit 0-3) select the variable number. Bits 4-7 select the HART slave address. If invalid values are
used, the temperature value will be invalid, with no Error bit set.
For example, if we want to use FV from HART Slave 3 as temperature source, we have to write the value 34 Hex (52 decimal).
eeee
An error value.
Bit 0: Invalid SV value (Float 0).
Bit 8: Invalid Temperature value.
Bit 12: Invalid PV value (Float 1).
Wwww A warning value, not used in this implementation.
Parameter Default value Configurable values
Baud Rate 9600 1200, 2400, 4800, 9600, 19200
Start Bits One One
Data Bits Seven Seven, Eight
Parity None None, Odd, even
Stop Bits One One or Two
Address 1 1-255